Low-Level Radon Activity Concentration—A MetroRADON International Intercomparison
Otahal, Petr P. S.; Fialova, Eliska; Vosahlik, Josef; Wiedner, Hannah; Grossi, Claudia; Vargas, Arturo; Michielsen, Nathalie; Turtiainen, Tuukka; Luca, Aurelian; Wołoszczuk, Katarzyna; Beck, Thomas R. (2022-05-10)
Tätä artikkelia/julkaisua ei ole tallennettu Julkariin. Julkaisun tiedoissa voi kuitenkin olla linkki toisaalle tallennettuun artikkeliin/julkaisuun.
Otahal, Petr P. S.
Fialova, Eliska
Vosahlik, Josef
Wiedner, Hannah
Grossi, Claudia
Vargas, Arturo
Michielsen, Nathalie
Turtiainen, Tuukka
Luca, Aurelian
Wołoszczuk, Katarzyna
Beck, Thomas R.
10.05.2022
This publication is copyrighted. You may download, display and print it for Your own personal use. Commercial use is prohibited.
International Journal of Environmental Research and Public Health : 10
Tiivistelmä
An international comparison of continuous monitors measuring radon activity concentration was performed to validate the traceability of the European radon calibration facilities. It was carried out by comparing the secondary standards used by these previous facilities, ranging from 100 Bq·m−3 to 300 Bq·m−3. Secondary standards were individually compared to a secondary reference device previously calibrated in a reference radon atmosphere traceable to a primary standard. The intercomparison was organized by the National Institute for Nuclear, Chemical, and Biological Protection (SUJCHBO) in the period from October 2019 to April 2020 within the European Metrology Program for Innovation and Research (EMPIR), JRP-Contract 16ENV10 MetroRADON. Eight European laboratories participated in this study. The results of the experiment are presented and discussed.